Abstract

The analysis of S-carboxymethyl-L-cysteine S-oxygenase activity by descending paper chromatography and chloroplatinate visualization is a cost effective but time consuming analytical procedure with 20 enzyme assays taking 88 h to analyse. Here we report on faster and more sensitive method of analysing S-carboxymethyl-L-cysteine S-oxygenase activity using high-pressure liquid chromatography with fluorescence detection following pre-column derivatization with o-phthaldehyde/2-mercaptoethanol. A base line separation of both S-carboxymethyl-L-cysteine R and S S-oxides was achieved using a 5μ, 30 × 2.1 mm id Hypersil Amino Acid C18 column with a gradient elution. The method was linear over the concentration range 2.5–250 μM for both S-carboxymethyl-L-cysteine R and S S-oxides. The total analysis time for the 20 enzyme assays was reduced from 88 h to11 h 40 min.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.