Abstract

The indocyanine green (ICG) finger-piece method (FPM), which allows measurement of the ICG concentration by mounting a light sensor onto a finger, is used to assess liver function. We compared the ICG FPM with the conventional ICG blood sampling method (BSM) in patients with liver disorders. Ninety consecutive patients simultaneously underwent the ICG BSM and ICG FPM. After ICG administration, blood samples were collected at 5, 10, and 15min for the ICG BSM. The ICG concentration was measured through the finger piece by an ICG clearance meter. Seventy-one patients (78.9%) had Child-Pugh class A liver disease, and 19 (21.1%) had class B or C. The FPM-measured ICG plasma disappearance rate was positively correlated with the BSM-measured values (r = 0.886, P <0.001). Bland-Altman analysis showed good agreement between the two methods (mean difference, 0.012 ± 0.018). The FPM-measured ICG plasma disappearance rate was positively correlated with the BSM-measured values both in patients with Child-Pugh class A liver disease (r = 0.821, P <0.001) and class B or C liver disease (r = 0.859, P <0.001). The ICG FPM may be an alternative to the ICG BSM for liver function assessment.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.