Abstract
Eosin has been used to generate the background signal for indirect fluorimetric detection of inorganic and organic ions, simultaneously separated by capillary zone electrophoresis (CZE). This reagent provides constant fluorescence over the pH range of 5–10 and is compatible with the excitation by an argon ion laser at 488 nm with emission at 520 nm. The use of esosine as fluorophore, H 3BO 3, and Na 2B 4O 7 as electrolyte and diethylentriamine as modifier of the electroosmotic flow in CZE were optimised. The analytical potential of the studied buffer was tested on a group of 12 anions, used as model compounds. Both, hydrodynamic and electrokinetic injection mode were optimised. The detection limits determined by the last injection mode, were in the range 0.008–0.037 mg l −1. By using this method, the quantitation of the common anions in tap and mineral water has been carried out successfully.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.