Abstract
A direct-current argon plasma spectrometer has been interfaced with a gel filtration chromatography column to serve as a multi-element-specific detector. This analytical system was used to speciate protein-bound copper, iron, and zinc in serum and intravenous infusion fluids. The operating parameters of the direct current argon plasma including instrumental drift, detection limits, effect of background levels on the calibration graphs. and accuracy were optimized. Calibrations had to be repeated every hour to compensate for instrumental drift. The detection limits of this system (3.2, 3.9 and 9.3 μg l −1 for copper, iron and zinc, respectively) are adequate for the determination of most species containing those elements in the column effluent.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.