Abstract

Background/purposeSegmental body defects of the mandible result in the complete loss of the affected region. In our previous study, we investigated the clinical applicability of a customized mandible prosthesis (CMP) with a pressure-reducing device (PRD) in an animal study. In this study, we further incorporated dental implants into the CMP and explored the use of dental implant PRD (iPRD) designs. Materials and methodsBy employing a finite element analysis approach, we created 4 types of CMP: CMP, CMP with iPRD, CMP-PRD, and CMP-PRD with iPRD. We developed 2 parameters for the iPRD: cone length (CL) in the upper part and spring pitch (SP) in the lower part. Using the response surface methodology (RSM), we determined the most suitable structural assignment for the iPRD. ResultsOur results indicate that CMP-PRD had the highest von Mises stress value for the entire assembly (1076.26 MPa). For retentive screws and abutments, CMP with iPRD had the highest von Mises stress value (319.97 and 452.78 MPa, respectively). CMP-PRD had the highest principal stress (131.66 MPa) in the anterior mandible. The iPRD reduced principal stress in both the anterior and posterior mandible. Using the RSM, we generated 25 groups for comparison to achieve the most favorable results for the iPRD and we might suggest the CL to 12 mm and the SP to 0.4 mm in the further clinical trials. ConclusionUse of the PRD and iPRD in CMP may resolve the challenges associated with CMP, thereby promoting its usage in clinical practice.