Abstract

The coronavirus disease 2019 (COVID-19) pandemic has created a need for remote blood glucose (BG) monitoring in the intensive care unit (ICU). To evaluate feasibility and patient safety of a hybrid monitoring strategy of point-of-care (POC) BG plus continuous glucose monitor (CGM) in the ICU. Retrospective analysis. ICU of an academic medical center. Patients with COVID-19 on IV insulin. After meeting initial validation criteria, CGM was used for IV insulin titration and POC BG was performed every 6 hours or as needed. Outcomes included frequency of POC BG, workflow, safety, and accuracy measures. The study included 19 patients, 18 with CGM data, mean age 58 years, 89% on mechanical ventilation, 37% on vasopressors, and 42% on dialysis. The median time to CGM validation was 137 minutes (interquartile range [IQR] 114-206). During IV insulin, the median number of POC values was 7 (IQR 6-16) on day 1, and declined slightly thereafter (71% reduction compared with standard of 24/day). The median number of CGM values used nonadjunctively to titrate IV insulin was 11.5 (IQR 0, 15) on day 1 and increased thereafter. Time in range 70 to 180 mg/dL was 64 ± 23% on day 1 and 72 ± 16% on days 2 through 7, whereas time <70 mg/dL was 1.5 ± 4.1% on day 1 and <1% on days 2 through 7. This study provides data to support that CGM using a hybrid protocol is feasible, accurate, safe, and has potential to reduce nursing and staff workload.