Abstract

Although diffusion tensor imaging (DTI) shows great potential for the diagnosis of a variety of pathologies, no consensus for an appropriate assessment standard of DTI exists. This study examined the feasibility of using water-filled arrays of glass capillaries to construct a DTI phantom suitable for making repeated and reproducible measurements required in a quality assessment program. Three phantoms were constructed using arrays of capillaries with three inner diameters (23, 48, and 82 microm). Data were acquired using DTI protocols; the fractional anisotropy (FA), mean apparent diffusion coefficient (ADC) and principal eigenvectors of the diffusion tensors were calculated. This study demonstrated four results: (1) echo-planar images show that susceptibility within the capillary arrays does not lead to substantial differences in precessional frequency in regions containing the arrays and neither do the regions show noticeable image distortion; (2) principal eigenvectors of the diffusion tensors agree to within<10.3 degrees of the array orientations; (3) mean FA values (0.18-0.50) and ADC values (1.40-1.93x10-(3) mm2/s) within specified regions of interest are in general agreement with simulations after a simple noise correction; and (4) these array performance characteristics are observable using a typical clinical DTI protocol.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.