Abstract

The accuracy of a digital impression technique to fabricate the implant restoration and abutment for a dental implant using an intraoral laser scanner was evaluated in 36 patients who were missing a single posterior tooth in either the mandible or maxilla that was restored with a single implant. The spatial position of each integrated implant, including the surrounding anatomic hard and soft tissues of adjacent structures, was captured utilizing a special scanning abutment with an intraoral laser scanner. Data from the scanning protocol was then delivered via the Internet in the form of an STL file to the manufacturing site for the production of a custom computer-aided design abutment and crown. All 36 restorations and abutments were delivered to the patients and evaluated for marginal integrity, interproximal contact points, and occlusion. Of the 36 patients, 6 required contact adjustments, 7 required occlusal adjustments, and 3 required a gingivectomy around the implant to completely seat the restoration. Chair time for adjustments did not exceed 15 minutes. The findings suggest that an intraoral laser scanner can be used with confidence to obtain consistent and accurate digital impressions to fabricate custom restorations and abutments for dental implants.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call