Abstract

The use of absorbents as retaining materials in the through oven transfer adsorption desorption interface (TOTAD) of an on-line coupled reversed-phase liquid chromatography–gas chromatography system (RPLC–GC) is proposed for the first time. A comparative study of an adsorbent (Tenax TA) and two absorbents, namely polydimethylsiloxane and poly(50% phenyl/50% methylsiloxane) is performed to establish the best experimental conditions for the automated and simultaneous determination of 15 organophosphorus and organochlorine pesticide residues in olive oil. The proposed method provides satisfactory repeatability (RSDs lower, in general, than 8.5%) and sensitivity (limits of detection ranging from 0.6 to 81.9 μg/L) for the investigated compounds.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.