Abstract
During transapical transcatheter aortic valve replacement (TA-TAVR), the apical closure remains a challenge for the surgeon, having the risk for ventricular tear and massive bleeding. Apical closure devices are already under clinical evaluation, but only a few can lead to a full percutaneous TA-TAVR. We describe the successful use of a 9-mm myocardial occluder (ventricular septal defect occluder) that was used to seal the apex after a standard TA-TAVR (using the Sapien XT 23-mm transcatheter valve and the Ascendra + delivery system). The placement of the nonmodified myocardial occluder was performed through the Ascendra + delivery system, with a very small amount of blood loss and an acceptable sealing of the apical tear. This approach is feasible and represents a further step toward true-percutaneous transapical heart valve procedures. Modified apical occluders are under evaluation in animal models.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.