Abstract
We sought to replace the highly hemolysis-susceptible diazo conjugated bilirubin (Bc) assay with the more robust vanadate oxidation method and determine its impact on test cancellation in the pediatric population. Analytical validation of the Randox vanadate assay and comparison with the Roche diazo method were performed. The frequency of pediatric sample cancellation because of hemolysis was compared between the diazo and vanadate methods by retrospective analysis of clinical test data. The vanadate assay demonstrated no clinically significant interference from hemolysis up to a hemolysis index of 1,300 (approximately 13 g/L hemoglobin). There was a strong correlation with the diazo method (r2 = 0.97) but with a positive slope bias of 1.27. Implementing the vanadate method resulted in a significantly lower proportion of pediatric samples cancelled because of hemolysis compared with the diazo method (0.6% of 688 patients vs 30.6% of 10,464 patients, respectively; P < .001), with a 0.6% (n = 513) vs 43.2% (n = 6,464) reduction in test cancellations (P < .001) for children younger than 6 months of age. The vanadate method showed robust performance against hemolysis. Its implementation resulted in a significant decrease in pediatric tests cancelled because of hemolysis.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.