Abstract

AbstractA system is described that allows the introduction of large volumes of water samples in capillary GC. Water elimination is carried out in the solvent split mode in a PTV injector with a packed liner. Two ways of separating water and analytes, i.e. evaporative and non‐evaporative (solid‐phase extraction), are compared. Sampling in the solid‐phase extraction mode is favorable both in terms of recovery as well as with regard to sampling time. Quantitative recovery is obtained for priority pollutants ranging in volatility from dimethyl‐phenol to phenanthrene. Losses occur for more volatile compounds, but even for these compounds the repeatability of the recoveries remains acceptable. With the system described here, water samples up to at least 1 ml of water can be directly analyzed. The detection limits are in the sub‐ppb range.