Abstract

To evaluate whether measurement of the hemoglobin (Hb) concentration with a blood gas analyzer approximates that determined by a conventional coulter counter in critically ill adults. Prospective patient series. A 32-bed cardiovascular, neurosurgical, trauma and medical-surgical intensive care unit in a single Canadian center. We consecutively recruited 202 critically ill adults, the majority of whom had recent cardiac or vascular surgery, neurosurgery or trauma. The nurse obtained a single arterial blood sample within a few hours of the patient's admission to the intensive care unit. The Hb concentration was determined from each blood sample in a masked fashion, using both a blood gas analyzer and a conventional laboratory coulter counter. A total of 202 consecutive paired analyses were conducted. There was a highly significant correlation between the coulter counter and blood gas analyzer methods of Hb measurement (r2 = 0.98, 95% confidence interval [CI] = 0.97-0.99; P < 0.0001). Using the method of Bland and Altman, the overall mean difference in Hb concentration between the coulter counter and the blood gas analyzer was -4.3 g/l (95% CI = -11.0 to 2.4). Of the 11 (5.4%) Hb measurements that extended beyond the upper and lower 95% CI, 10 (5.0%) were within +/- 3 g/l of these confidence limits. An arterial blood gas analyzer may provide a valid alternative method to the traditional coulter counter for the rapid assessment of Hb concentration among critically ill adults. Since issues related to its safety, quality control, data entry and cost savings have yet to be addressed, however, use of such point of care testing should be viewed as a supplement to conventional laboratory testing.

Highlights

  • The frequent collection of blood specimens in the intensive care unit (ICU) may contribute to iatrogenic blood loss [1,2], compounding the problem of acute anemia that is frequently found among individuals with active hemorrhage, with a bleeding diathesis or with hemodilutional anemia [3]

  • Because little is known about the accuracy of arterial blood gas (ABG) analysis for the determination of Hb concentration, we prospectively evaluated its use in a critical care setting

  • The mean Hb concentration measured by the coulter counter was 102.9 g/l, while that measured by the ABG was 107.2 g/l

Read more

Summary

Introduction

The frequent collection of blood specimens in the intensive care unit (ICU) may contribute to iatrogenic blood loss [1,2], compounding the problem of acute anemia that is frequently found among individuals with active hemorrhage, with a bleeding diathesis or with hemodilutional anemia [3]. The lag time in the diagnosis of severe acute anemia, due to long laboratory turnaround times, may have important clinical consequence [2,4,5]. The availability of arterial blood gas (ABG) analyzers in most. ICUs and hospital laboratories enables rapid analysis of traditional blood gas elements, such as pH and pO2, and Hb concentration. Because little is known about the accuracy of ABG analysis for the determination of Hb concentration, we prospectively evaluated its use in a critical care setting

Methods
Results
Discussion
Conclusion