Abstract

It is well known that photomultipliers (PMTs) are sensitive to external magnetic fields. We have used this property to vary the gain of the PMT depending on the position of the light source. Various coil configurations have been evaluated to produce local magnetic fields in the region between the photocathode and first dynode to maximize the amplitude variation in the anode signal with respect to change in the coil current. The aim was to improve the position sensitivity of a PMT used in positron emission tomography (PET). The position sensitivity was tested with an array of collimated light emitting diodes directed towards the photocathodes of a Hamamatsu R1548 PMT. The best coil position was above the photocathodes. A 60% reduction in output for light pulses from two LEDs near the coil was obtained with a current of 43 Ampere-turns while signals from two LEDs far from the coil remained the same. The rise time of the magnetic field was 2 μs.