Abstract

Quality assurance methods used in association with radiosurgery must include all aspects of the radiosurgery process: visualization and localization of the target, treatment and dose planning and dose delivery. Presented here is a quality assurance method that utilizes an anthromorphic head phantom and a micro-ionization chamber to demonstrate precise target localization and accurate dose delivery. This micro-ionization chamber method offers an immediate readout which is both accurate and reproducible. Additionally, this method allows unlimited repetition of the dose measurement process without repeated radiographic localization studies as is necessary with the conventional methods of TLD, film, and Fricke gels. The method and techniques presented can be used in the acceptance testing and routine quality assurance of both linac-based and Gamma Knife radiosurgery units.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.