Abstract
To evaluate the clinical usefulness of a stent retrieval hook in the removal of occluded lacrimal stents. The hook consisted of an inner hook part and an outer pusher part. Under administration of local anesthetic, the retrieval hook was introduced into the nasal cavity and aimed laterally toward the inferior meatus to grasp the distal end of the stent. After grasping the end of the stent, the sheath was pushed to fully grasp it. After that, the hook was pulled out of the external naris. The retrieval hook was used to remove 267 occluded lacrimal plastic or metallic stents. Technical success rate was 97% in the removal of plastic stents and 96% in the removal of metallic stents. The causes of technical failure were entrenchment of granulation tissue into the stent in six plastic stents and one metallic stent, and inaccessibility of the hook to the distal tip of the stent in one plastic stent. These eight stents were removed by an otorhinolaryngologist or an ophthalmologist. Epistaxis occurred in six patients, which was controlled by a nasal pack. The lacrimal stent retrieval hook seems useful for the removal of occluded stents.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.