Abstract
This Project evaluated a human visual system model (JNDmetrix) based on just noticeable difference (JND) and frequency-channel vision-modeling principles to assess whether a Cathode ray tube (CRT) or a liquid crystal display (LCD) monochrome display monitor would yield better observer performance in radiographic interpretation. Key physical characteristics, such as veiling glare and modulation transfer function (MTF) of the CRT and LCD were measured. Regions of interest from mammographic images with masses of different contrast levels were shown once on each display to six radiologists using a counterbalanced presentation order. The images were analyzed using the JNDmetrix model. Performance as measured by receiver operating characteristic (ROC) analysis was significantly better overall on the LCD display (P = 0.0120). The JNDmetrix model predicted the result (P = 0.0046) and correlation between human and computer observers was high (r (2) (quadratic) = 0.997). The results suggest that observer performance with LCD displays is superior to CRT viewing, at least for on-axis viewing.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.