Abstract

In capillary column liquid chromatography, a cylindrical flow-through cell is often used in absorbance detection. The cross-sectional shape of the cylindrical flow-through cell perpendicular to the light beam is rectangular. Therefore, the optimum shape of the focused light beam is also rectangular. The use of a cylindrical lens to obtain a rectangular light beam, thus increasing the sensitivity of the UV detector more than 10-fold, is described.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.