Abstract
The purpose of this study was to assess the usefulness of a cauterizing laparoscopic linear stapler for intestinal anastomosis. In a porcine model, intestinal anastomoses performed with a standard laparoscopic linear stapler, a cauterizing laparoscopic linear stapler (RF stapler), and a two-layer, hand-sewn technique were compared by measuring bursting pressures at 4 and 7 days after surgery. During surgery, the RF stapler provided better hemostasis than the regular stapler for mesenteric transection. At 4 days, one leak occurred in the RF stapler group, and the bursting pressure in the RF stapler group was significantly lower than the bursting pressures in the regular stapler group and the hand-sewn group. In addition, the bursting pressure was significantly greater in the hand-sewn group than in the regular stapler group at 4 days. By 7 days, there were no differences in bursting pressure among the groups. We recommend that the RF stapler not be used for intestinal anastomosis. However, the device may be beneficial for controlling vasculature.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Surgical Laparoscopy, Endoscopy & Percutaneous Techniques
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.