Abstract

The use of a diabetic blood glucose meter for radiochromic film dosimetry in blood irradiation using x-ray beams on a medical linear accelerator has been investigated. The glucose meter provides optical density analysis in the visible and infrared region using a reflectance measurement technique. By comparing the ‘blood sugar’ level output with standard calibration gafchromic films a calibration curve is produced for quantitative analysis. Results show that a reproducible dose to meter output curve can be fitted using a second order polynomial function and that blood irradiation doses in vitro were measured to within 7.9% mean error (as compared to ionization chamber results) using the blood glucose meter. This level of accuracy falls below that measured with a standard densitometer (4.3%); however, results show that the blood glucose meter, which would be available in any haematology department, produces an adequate measure of gafchromic film optical density for blood irradiation dosimetry.