Abstract

Accurate registration of computed tomography (CT) images to the patient is crucial for computer-assisted surgery. Markers used in roentgen stereo-grammetric analysis (RSA) can be located in a CT scan using a novel approach and also can be located physically. Roentgen stereo-grammetric analysis data act as "ground truth" for the three-dimensional marker locations. Two foam-bone phantoms were marked. The markers were scanned seven times with RSA, three times with axial CT, and contacted four times with a coordinate measuring arm. Root-mean-square (RMS) errors were derived for the registrations. Computed tomography and RSA data register to 0.15 mm RMS error. Computed tomography and arm data register to 0.25 mm. The markers are biocompatible, and the coordinate measuring arm is usable in an operating room. Typical in vitro registration errors are approximately 2 mm. The authors have developed a marker that provides superior registration, and a procedure that can be used for in vivo studies.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.