This study shows the method and result on usability testing of ultrasonic surgical system for general liposuction & lower eyelid surgery. The purpose of the study is to validate whether UI risk mitigation controls are properly established and collected any opinions about the device. The quantitative data(task completion rate, the number of use error occurred, user satisfaction score) and qualitative data(subjective opinions via follow-up question) were collected. Due to simple how to use, few use errors were occurred. But the result from analyzing subjective data was different, several potential harms causing from the user interface were expected. These test results could help medical device designers to understand human factor considerations.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Proceedings of the International Symposium on Human Factors and Ergonomics in Health Care
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.