Abstract

This article reports on the development of usability engineering recommendations for next-generation integrated interoperable medical devices. A model-based hazard analysis method is used to reason about possible design anomalies in interoperability functions that could lead to use errors. Design recommendations are identified that can mitigate design problems. An example application of the method is presented based on an integrated medical system prototype for postoperative care. The AAMI/UL technical committee used the results of the described analysis to inform the creation of the Interoperability Usability Concepts, Annex J, which is included in the first edition of the new ANSI/AAMI/UL 2800-1:2019 standard on medical device interoperability. The presented work is valuable to experts developing future revisions of the interoperability standard, as it documents key aspects of the analysis method used to create part of the standard. The contribution is also valuable to manufacturers, as it demonstrates how to perform a model-based analysis of use-related aspects of a medical system at the early stages of development, when a concrete implementation of the system is not yet available.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.