Abstract

Background and aimsWe assessed the possibility to rule out negative urine cultures by counting with UriSed 3 PRO (77 Elektronika, Hungary) at Helsinki and Uusimaa Hospital District. Materials and methodsBacteria counting of the UriSed 3 PRO automated microscope was verified with reference phase contrast microscopy against growth in culture. After acceptance into routine, results of bacteria and leukocyte counting from 56 426 specimens with eight UriSed 3 PRO instruments were compared against results from parallel samples cultured on chromogenic agar. Laboratory data including preanalytical details were accessed through the regional database of the Helsinki and Uusimaa Hospital District. ResultsA combined sensitivity of 87–92% and a negative predictive value of 90–96% with a specificity of 54–50% was reached, depending on criteria. Preanalytical data (incubation time in bladder) combined with the way of urine collection would improve these figures if reliable. ConclusionsComplex patient populations, regional logistics and data interfases, and economics related to increased costs of additional particle counts against costs of screening cultures of all samples, did not support adaptation of a screening process of urine cultures. This conclusion was made locally, and may not be valid elsewhere.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call