Abstract

A sensitive and specific ion-pair reversed-phase high performance liquid chromatography (HPLC) method for urinary iodine analysis is described. This method is based on pulsed amperometric detection (PAD) using a silver working electrode (HPLC–PAD), which improves peak shape, electrode stability as well as linearity and reproducibility. A two-step extraction process consisting of solid phase extraction (SPE) and liquid–liquid extraction with dichloromethane was added in order to improve sample purification which is essential with the use of PAD. Treated samples were eluted on a C18 column, using a phosphate buffer containing ion-pairing reagent tetrabutylammonium and 5% MeOH. The calibration standard curves were linear up to 500μg/L and within-run and between-run coefficients of variation (CVs) were <6% with the quantification limit fixed at 6μg/L. Accuracy, expressed as recovery, ranged from 94% to 104%. Comparison with the Technicon AutoAnalyzer acid digestion (AA) method resulted in a high correlation (r=0.9916). Due to a low quantification limit and high sample throughput, the proposed technique appears suitable for both epidemiological and clinical follow-up studies.