Abstract

To measure low neonatal gonadotropin levels, a sensitive non-invasive method is optimal. The aim of the current study was to validate the Architect i2000SR, an automated immunoassay analyser for the measurement of gonadotropins in unextracted neonatal urine samples against serum gonadotropin levels as a gold standard. Blood and urine were sampled from 30 approximately six-week-old male and female neonates undergoing elective paediatric surgery. Follicle stimulating hormone (FSH) and luteinizing hormone (LH) were measured and the urine results were corrected for creatinine. The agreement between neonatal serum and urinary FSH was 0.904 (3-5 h between samples) and 0.704 (18-20 h). For LH, the correlation coefficients were 0.785 and 0.507, respectively. We conclude that gonadotropins can be reliably measured using the Architect on randomly voided, non-extracted urine samples collected from neonates by an adhesive device. Urinary gonadotropin levels are a proper reflection of the serum levels.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.