Abstract
The development of a new electrochemical sensor consisting in a glass-sealed metal microelectrode coated by a polyethylenimine film is described. The use of polymers as the entrapping matrix for enzymes fulfils all the requirements expected for these materials without damaging the biological material. Since enzyme immobilization plays a fundamental role in the performance characteristics of enzymatic biosensors, we have tested four different protocols for enzyme immobilization to determine the most reliable one. Thus the characteristics of the potentiometric biosensors assembled were studied and compared and it appeared that the immobilization method leading to the most efficient biosensors was the one consisting in a physical adsorption followed by reticulation with dilute aqueous glutaraldehyde solutions. Indeed, the glutaraldehyde immobilized urease sensor provides many advantages, compared to the other types of sensors, since this type of urea biosensor exhibits short response times (15–30 s), sigmoidal responses for the urea concentration working range from 1×10 −2.5 to 1×10 −1.5 M and a lifetime of 4 weeks.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.