Abstract

Uranium single particle analysis has been performed by inductively coupled plasma-mass spectrometry (ICP-MS) and the performances are compared with that provided by scanning electron microsopy and single particle counting. The transient signal induced by the flash of ions due to the ionisation of an uranium colloidal particle in the plasma torch can be detected and measured for selected uranium ion masses ( 238U +, 235U + or 254[ 238U 16O] +) by the mass spectrometer. The signals recorded via time scanning are analysed as a function of particle size or fraction of the studied element or isotope in the colloid phase. The frequency of the flashes is directly proportional to the concentration of particles in the colloidal suspension. The feasibility tests were performed on uranium dioxide particles. The study also describes the experimental conditions and the choice of mass to detect uranium colloids in a single particle analysis mode.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.