Abstract

Transmission of oesophageal images may vary between different small-bowel capsule endoscopy models. A retrospective review of 100 examinations performed with 2 different Small-bowel capsule endoscopy (SBCE) systems (PillCam(®) and MiroCam(®)) was performed. The oral cavity/aero-digestive tract (i.e., tongue, uvula and/or epiglottis) was captured/identified in almost all (99%) of PillCam(®) videos but in none of MiroCam(®) cases, P < 0.0001. Furthermore, oesophageal images (i.e., from the upper oesophageal sphincter to the Z-line were captured in 99% of PillCam(®) videos (mean ± SD, 60.5 ± 334.1 frames, range: 0-3329 frames) and in 66% of MiroCam(®) cases (mean ± SD, 11.1 ± 46.5 frames, range: 0-382 frames), P < 0.0001. The Z-line was identified in 42% of PillCam(®) videos and 17% of MiroCam(®), P = 0.0002. This information might be useful when performing SBCE in patients with high risks for aspiration.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.