Abstract

A simultaneous ultra-performance liquid chromatography photodiode array detection method was developed and validated for Elvitegravir, Tenofovir and Emtricitabine and Cobicistat drugs in tablet dosage forms. Effective liquid chromatographic quantification was achieved using a Luna C18 (100 × 2.6mm, 1.6μ) column with an acetonitrile: 0.1% v/v trifluoro acetic acid (80:20% v/v) mobile phase and a photo diode array detector at 260nm for a shorter run time of 3min. The respective retention times for Elvitegravir, Tenofovir, Emtricitabine and Cobicistat were 0.311, 0.792, 1.379 and 2.045min. The range of linearity was 37.50-225.00, 2.50-15.00, 50.00-300.00and 37.50-225.00μg/mL for Elvitegravir, Tenofovir, Emtricitabine and Cobicistat, respectively. A mass spectrometer was employed for quantitative analysis using electron spray ionization multiple reaction monitoring in positive mode. The developed method was validated for different parameters to meet the acceptance criteria as per ICH Q2 (R1) guidelines. Using UPLC-ESI-MS, 12 degradants were identified and their mechanisms were studied.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.