Abstract

We present in this contribution the new system for Laser calibration of the ATLAS hadronic calorimeter TileCal. The Laser system is a part of the three stage calibration apparatus designed to compute the calibration factors of the individual cells of TileCal. The Laser system is mainly used to correct for short term drifts of the readout of the individual cells. A sub-percent accuracy in the control of the calibration factors is required. To achieve this goal in the LHC Run2 conditions, a new Laser system was designed. The architecture of the system is described with details on the new optical line used to distribute Laser pulses in each individual detector module and on the new electronics used to drive the Laser, to read out optical monitors and to interface the system with the ATLAS readout, trigger and slow control. The LaserII system has been fully integrated into the framework used for measuring calibration factors and for monitoring data quality. First results on the Laser system performances studied are presented.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call