Abstract

A major upgrade has been implemented for the ns-gated laser entrance hole imager on the National Ignition Facility (NIF) to obtain high-quality data for Hohlraum physics study. In this upgrade, the single "Furi" hCMOS sensor (1024 × 448 pixel arrays with two-frame capability) is replaced with dual "Icarus" sensors (1024 × 512 pixel arrays with four-frame capability). Both types of sensors were developed by Sandia National Laboratories for high energy density physics experiments. With the new Icarus sensors, the new diagnostic provides twice the detection area with improved uniformity, wider temporal coverage, flexible timing setup, and greater sensitivity to soft x rays (<2 keV). These features, together with the fact that the diagnostic is radiation hardened and can be operated on the NIF for high neutron yield deuterium-triterium experiments, enable significantly greater return of data per experiment.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call