Abstract
After dialysis ends, urea continued movement causes rebound postdialysis, with values at about 20%. New techniques have been incorporated into hemodialysis, but their relationship with rebound has not yet been studied. This study aimed to quantify urea rebound at 30-min postdialysis during sessions using polysulfone filters and high-flow versus online hemodiafiltration, and to define its correlation with body composition measured by bioimpedance by a cross-sectional study with 69 patients (December 2015 to January 2016). Mean urea rebound was 24.39, which was positively associated with recirculation, Kt/V or hypotension, and showed a negative relationship with online hemodiafiltration. It was not associated with different body composition compartments. To conclude, postdialysis urea rebound remained high with polysulfone dialyzers and low dialysis doses. Online hemodiafiltration could improve postdialysis urea rebound. Different body composition compartments were not related to rebound.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Therapeutic apheresis and dialysis : official peer-reviewed journal of the International Society for Apheresis, the Japanese Society for Apheresis, the Japanese Society for Dialysis Therapy
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.