Since 2009, three national metrology institutes (NMI) have submitted three samples of known activity of 241Am to the International Reference System (SIR) for activity comparison at the Bureau International des Poids et Mesures (BIPM), with comparison identifier BIPM.RI(II)-K1.Am-241. The values of the activity submitted were between about 3 MBq and 38 MBq. These three results replace their earlier results obtained in the linked CCRI(II)-K2.Am-241 in 2003. There are now five results in the BIPM.RI(II)-K1.Am-241 comparison, 16 results remaining from the CCRI(II)-K2.Am-241 comparison and two results in the COOMET.RI(II)-K2.Am-241 comparison. The degrees of equivalence between each equivalent activity measured in the SIR and the KCRV dating 2007 have been calculated and the results are given in the form of a table. A graphical presentation is also given.Main textTo reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.The final report has been peer-reviewed and approved for publication by the CCRI, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.