Gradient elution of polystyrene standards on reversed phase C18 columns by methanol/tetrahydrofuran or methanol/dichloromethane mixtures yielded a strange effect of the molecular weight of the sample on the specific peak are (mAUs per μg sample injected). The effect did not occur when pure dichloromethane was used as an eluent. Further to this, dependence of specific peak area on flow rate was observed in gradient elution with methanol/tetrahydrofuran mixtures. It was found that these effects were due to polymer elution at the verge of precipitation. Depending on the dwell time of the sample in the column, opalescence occurred that added to the UV signal used for monitoring the elution.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.