Abstract

We report a case of 37-year-old man implanted with cardiac resynchronization therapy-defibrillator presented with persistent low-grade fever and sudden loss of left ventricular (LV) capture from coronary sinus lead after generator replacement. 18 F-fluorodeoxyglucose positron emission tomography with computed tomography scan showed increased uptake at posterolateral region of the pericardium adjacent to the LV lead, suggestive of possible lead-related infection. Combined percutaneous and surgical lead extraction revealed purulent pericarditis and polymerase chain reaction testing confirmed tuberculous (TB) pericarditis. TB pericarditis is an unusual cause of loss of LV capture, but should be considered in countries where TB is still endemic.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.