Abstract

The objective of our study was to illustrate a case of endoscopically placed biliary stent breakage. A72-year-old woman with a prolonged history of cholangitis following laparoscopic cholecistectomy was referred to our institution 8 years ago. Dilatation of the intra- and extrahepatic biliary tree and a benign stricture at the cystic confluence were observed at US and endoscopic retrograde cholangiopancreatography (ERCP). A 12-F gauge plastic endoprosthesis was placed. In the absence of any symptoms, breakage of the stent was revealed 18 months later at plain radiology. Eight years later an enterocutaneous fistula occurred originating from a jejunal loop containing the indwelled distal part of the stent. Surgery was undertaken and the distal part of the stent removed with the perforated jejunal loop. The proximal part was successively endoscopically removed. Disruption of a biliary endoprosthesis is observed in patients in whom the stent is kept in situ for a long period or consequent to exchange. The removal and exchange is mandatory when the stent disruption is followed by cholangitis. In the current case, because of the absence of any symptoms the removal of the stent was not attempted. Immediate endoscopic removal of the prosthetic fragments seems to be the treatment of choice for replacement of a new stent.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.