Abstract
This first part of the paper describes the calculation and design of a symmetric, double focusing 75° sector field spectrometer which is used for response measurements of solid state detectors given in part 2. The design feature was derived from the use as a versatile laboratory instrument for generating a monochromatic parallel beam of electrons of energies up to 3 MeV. The specific data of the instrument are: Radius of curvature: ϱ = 6.62 cm , Distance effective-field-boundary to radioactive source: l = 17.25 cm , Resolution: Δp/p = 1.5%, Solid angle: Ω = 0.1%, Pole-piece gap: k = 2.0 cm .
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.