Abstract
A newborn baby with persistent neonatal jaundice developed severe septicaemia. Blood glucose monitoring using Accuchek(R) revealed persistently high values, but laboratory blood glucose testing showed no instance of true hyperglycaemia. She was found to have galactosaemia. The Accu-chek® glucose meter does not discriminate between blood galactose and glucose.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.