Abstract

Nasal high flow (NHF) cannulae are used to deliver heated and humidified air to patients at steady flows ranging from 5 to 50 l/min. In this study, the flow velocities in the nasal cavity across the complete respiratory cycle during natural breathing and with NHF has been mapped in vitro using time-resolved stereoscopic particle image velocimetry (SPIV). An anatomically accurate silicone resin model of a complete human nasal cavity was constructed using CT scan data and rapid prototyping. Physiological breathing waveforms were reproduced in vitro using Reynolds and Womersley number matching and a piston pump driven by a ball screw and stepper motor. The flow pattern in the nasal cavity with NHF was found to differ significantly from natural breathing. Velocities of 2.4 and 3.3 ms−1 occurred in the nasal valve during natural breathing at peak expiration and inspiration, respectively; however, on expiration, the maximum velocity of 3.8 ms−1 occurred in the nasopharynx. At a cannula flow rate of 30 l/min, maximal velocities of 13.6 and 16.5 ms−1 at peak expiration and inspiration, respectively, were both located in the cannula jet within the nasal valve. Results are presented that suggest the quasi-steady flow assumption is invalid in the nasal cavity during natural breathing; however, it was valid with NHF. Cannula flow has been found to continuously flush the nasopharyngeal dead space, which may enhance carbon dioxide removal and increase oxygen fraction.