Abstract

Stent fracture is a rather infrequent complication associated with in-stent restenosis, thrombosis, aneurysm formation, and ischemic events. Several stent-related parameters, such as the use of longer or multiple stents, stent overlapping, and balloon/stent overexpansion are potential predictors of stent fracture. Stents deployed in right coronary artery lesions with exaggerated motion, tortuosity, or severe calcification are also generally considered to be at higher risk for fracture. This case demonstrates that intravascular ultrasound imaging is extremely useful to confirm the diagnosis and identify the possible mechanism of the stent fracture, as well as to assess the final result after subsequent angioplasty.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.