Abstract

Liquid 1-decanethiol was confined on an atomic force microscope (AFM) tip apex and the effect was investigated by measuring amplitude-distance curves in dynamic force mode. Within the working distance in the dynamic force mode AFM, the thiol showed strong interactions bridging between a gold-coated probe tip and a gold-coated Si substrate, resulting in unstable amplitude and noisy AFM images. We show that under such a situation, the amplitude change is dominated by the extra forces induced by the active material loaded on the tip apex, overwhelming the amplitude change caused by the geometry of the sample surface, thus resulting in noise in the image the tip collects. We also show that such a contaminant may be removed from the apex by pushing the tip into a material soft enough to avoid damage to the tip.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.