Abstract

The extent to which the diagnostic performance of low-dosed storage phosphors can be improved by suitable image postprocessing was investigated. Storage phosphor radiographs and conventional 200- and 400-speed screen-film combinations of an anthropomorphic chest phantom with simulated nodules and linear structures were compared. The detection of test bodies was determined in a ROC analysis based on 16200 individual observations. The two analog screen-film systems did not differ significantly. As far as the storage phosphor radiographs are concerned, a higher diagnostic performance was achieved with a large filter kernel (S 40) whereas small filter kernels and edge-enhanced imaging modes reduced the detection of detail. The image quality of low-dosed storage phosphors is thus similar to high-amplification screen-film combinations if large filter kernels are used in postprocessing of the image.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.