Abstract

Serial ostial and distal left main lesions continue to be one of the most difficult tasks for the interventional cardiologist, with many potential complications occurring. We present such a high-risk percutaneous coronary intervention where immediate stent explantation was deemed necessary because the metal deformation and high radial strength of the particular stent platform would prevent an acceptable procedural result if it had been crushed to the vessel wall. The aim of this paper was to discuss left main stent deformation, debate the risks and benefits of stent explantation and finally test in-vitro our theory on "insufficient" crush with stents with high radial strength and compare it with conventional stents. Bench-testing supports our ad-hoc explantation decision showing stent underexpansion, recoil, and malapposition, obtaining an inadequate minimal stent area.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.