Abstract

AbstractIt was the objective of this paper to assess the applicability of the universal calibration method to aqueous GPC/SEC with nonionic and anionic polymers using the Viscotek differential viscosity detector. Three water‐soluble polymers—polyacrylic acid, dextran, and polyethylene oxide—were chromatographed using four UltrahydrogelTM columns with 0.3 M NaCl and 0.1M KH2PO4 as the mobile phase adjusted to pH 7. Three distinct calibration curves were obtained. Upon addition of 10% methanol, a reasonably good universal calibration curve was obtained. However, quantitative analysis of the data exhibited about 5% deviation in average Mw and Mn for sodium polyacrylate as calculated from the single curve as opposed to about 40% when calculated from the composite curve. The applicability of three theoretical models for the universal calibration method was assessed, and a recommendation was made for future work. © 1993 John Wiley & Sons, Inc.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.