Abstract
A simple colorimetric flow-injection system for the determination of Cu(II) based on a complexation reaction with nitroso-R salt is described. The chemical and FIA variables were established using the univariate and simplex methods. A small volume of Cu(II) was mixed with merged streams of nitroso-R salt and acetate buffer solutions. The absorbance of the complex was continuously monitored at 492 nm. The calibration curve over the concentration range 1.0-7.0 microg ml(-1) was obtained. The relative standard deviation for determining 4.0 microg ml(-1) Cu(II) was 0.47% (n = 11). The detection limit (3sigma) was 0.68 microg ml(-1) and the sample throughput was 150 h(-1). The validity of the method has been satisfactorily examined for the determination of Cu(II) in wastewater and copper ore samples. The accuracy was found to be high, because the student t-values were calculated to be less than the theoretical values when the results were compared with those obtained by FAAS.