Abstract

Interruptions in continuous nebulized pulmonary vasodilators, such as epoprostenol, can potentially result in clinical deterioration in respiratory status. Coadministration of other intermittent nebulized therapies may require opening the ventilator circuit to facilitate administration. However, in patients with SARS-CoV2 infection, it is preferred to avoid opening the circuit whenever feasible to prevent aerosolization of the virus and exposure of health care workers. In this study, we describe a unique method of administering continuous epoprostenol nebulization and intermittent nebulized antibiotics, mucolytics, and bronchodilators, using Aerogen vibrating mesh nebulizers without interruptions in epoprostenol or opening the ventilator circuit. This technique set up consisted of stacking two Aerogen nebulizer cups, each with its own controller. This approach was successful in allowing concomitant delivery of intermittent and continuous nebulized therapy without interruptions. To our knowledge, this method has not been previously described in the literature and may be helpful to bedside clinicians facing a similar clinical scenario.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.