Abstract
Ricin, a toxic protein from the castor plant, is of forensic and biosecurity interest because of its high toxicity and common occurrence in crimes and attempted crimes. Qualitative methods to detect ricin are therefore needed. Untargeted liquid chromatography-tandem mass spectrometry (LC-MS/MS) proteomics methods are well suited because of their high specificity. Specificity in LC-MS/MS comes from both the LC and MS components. However, modern untargeted proteomics methods often use nanoflow LC, which has less reproducible retention times than standard-flow LC, making it challenging to use retention time as a point of identification in a forensic assay. We address this challenge by using retention times relative to a standard, namely, the uniformly 15N-labeled ricin A-chain produced recombinantly in a bacterial expression system. This material, added as an internal standard prior to trypsin digestion, produces a stable-isotope-labeled standard for every ricin tryptic peptide in the sample. We show that the MS signals for 15N and natural isotopic abundance ricin peptides are distinct, with mass shifts that correspond to the numbers of nitrogen atoms in each peptide or fragment. We also show that, as expected, labeled and unlabeled peptides coelute, with relative retention time differences of less than 0.2%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Similar Papers
More From: Analytical chemistry
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.