Abstract

In the study, the passive multi-layer neutron spectrometer, based on thermoluminescence detectors, was tested in a calibration laboratory with 239Pu-Be and 252Cf isotopic sources. MCNP code was used for the calculation of the response functions for the neutron energy range from 1 meV to 100MeV. It was also utilised for initial guess spectra calculations. Deconvolution was performed with MAXED and GRAVEL deconvolution codes resulting in the neutron spectra defined at the measuring point in the calibration laboratory.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call