Abstract
During a study with a scanning electron microscope to evaluate the structure of microspinal catheter after its removal from subarachnoid space, we found an unusual case. The observation with the microscope of the tip of a catheter removed at the end of an operation for hip replacement in a old female showed the presence of grounded particles with a crystal shape covering the outer surface. Further analysis of this material with an Energy-Dispersive Spectrometer (EDS) showed that it was barium. The patient performed a large bowel barium enema 8 months earlier for a painful syndrome to the lower abdomen. Authors rule out the contamination from the skin and suggest two possible mechanisms of passage of barium from blood to cerebrospinal fluid (CSF) and so to the surface of the catheter.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.